M5 Corpus Power Wheelchair REF,120032

Agency
U.S. Food and Drug Administration
Recall number
Z-2805-2026
Category
Medical devices
Recall date
2025-06-19
Enforcement report date
2026-08-05
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Permobil Inc

From the FDA Enforcement Report dated 2026-08-05. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
M5 Corpus Power Wheelchair REF,120032
Hazard / reason
This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.
Quantity
715 wheelchairs
Distribution
Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.
Code information
Power Platform firmware versions 1.27 through 1.57 UDI-DI 17330818334562
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 17330818334562

Brands