EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2810-2026
Category
Medical devices
Recall date
2026-06-24
Enforcement report date
2026-08-05
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medtronic Perfusion Systems

From the FDA Enforcement Report dated 2026-08-05. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78222
Hazard / reason
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Quantity
3,621 units
Distribution
Worldwide - US Nationwide distribution.
Code information
UDI-DI: 20763000135700, Lot Numbers: 0232602300 0232602383 0232612916 0232680078 0232708967 0232709001 0232910880 0232910881 0232910882 0233077983 0233245850 0233245860 0233245980; UDI-DI: 00763000135706, Lot Numbers: 0232612916 0232709001
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 20763000135700
  • 763000135706
Possible model numbers (as listed; not verified as model numbers)
  • 78222

Brands