EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2811-2026
- Category
- Medical devices
- Recall date
- 2026-06-24
- Enforcement report date
- 2026-08-05
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medtronic Perfusion Systems
From the FDA Enforcement Report dated 2026-08-05. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2811-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78322
- Hazard / reason
- Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
- Quantity
- 1,424 units
- Distribution
- Worldwide - US Nationwide distribution.
- Code information
- UDI-DI: 20199150023251, Lot Numbers: 0232709184 0232745036 0232798416 0232862784 0232862787; UDI-DI: 00199150023257, Lot Numbers: 0232862787
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 20199150023251
- 199150023257
- Possible model numbers (as listed; not verified as model numbers)
- 78322