OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2820-2026
- Category
- Medical devices
- Recall date
- 2026-07-02
- Enforcement report date
- 2026-08-05
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- IsoTis OrthoBiologics, Inc.
From the FDA Enforcement Report dated 2026-08-05. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2820-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095
- Hazard / reason
- Demineralized bone matrix (DBM) in resorbable mesh indicated for use with autograft as a bone graft extender in the posterolateral spine and pelvis, were filled with approx. 4.5 cc DBM (fill for 45 ¿ 11 mm configuration) instead of approx. 7.5 cc DBM (fill for 45 ¿ 17.5 mm configuration), which may lead to delayed fusion, incomplete fusion, pseudarthrosis, or lack of bone ingrowth.
- Quantity
- 98
- Distribution
- US Nationwide distribution in the states of AK, NJ, TN, AR, CA, WA, OH.
- Code information
- REF/UDI-DI/Lots: 56800095/10889981149352/181559, 181550 02-8000-095/10889981149314/181559, 181550
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II