Brand Name: Quantum Rehab Product Name: Q6 Edge 4 Model/Catalog Number: Q6 Edge 4 Software Version: 2.0.11, new…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2850-2026
Category
Medical devices
Recall date
2026-06-25
Enforcement report date
2026-08-12
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Pride Mobility Products Corp.

From the FDA Enforcement Report dated 2026-08-12. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: Quantum Rehab Product Name: Q6 Edge 4 Model/Catalog Number: Q6 Edge 4 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Hazard / reason
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Quantity
1102
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.
Code information
Lot Code: Q6 Edge 4 0060650920028*8 K143383 D246699 41875 ITI
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands