Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL

Agency
U.S. Food and Drug Administration
Recall number
Z-2880-2026
Category
Medical devices
Recall date
2026-06-26
Enforcement report date
2026-08-12
Classification
Class III
Status as reported by FDA
Ongoing
Manufacturer / firm
Siemens Healthcare Diagnostics, Inc.

From the FDA Enforcement Report dated 2026-08-12. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL
Hazard / reason
Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
Quantity
8 US; 44 OUS
Distribution
Worldwide - US Nationwide distribution in the states of CA and VA. The country of Japan.
Code information
UDI-DI 00842768001406 Lot U1
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class III

Identifiers

Barcodes (UPC/GTIN)
  • 842768001406

Brands