Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL

Agency
U.S. Food and Drug Administration
Recall number
Z-2881-2026
Category
Medical devices
Recall date
2026-06-26
Enforcement report date
2026-08-12
Classification
Class III
Status as reported by FDA
Ongoing
Manufacturer / firm
Siemens Healthcare Diagnostics, Inc.

From the FDA Enforcement Report dated 2026-08-12. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL
Hazard / reason
out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
Quantity
494 US; 114 OUS
Distribution
Worldwide - US Nationwide distribution in the states CA, ID, IL, IN, MA, MS, NC, NM, NY, OH, OK, OR, PA, WA, and Northern Mariana Islands part of USA and the countries of Canada, South Africa, Brazil, Chile, Colombia.
Code information
UDI-DI 00842768014321 Lot U2
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class III

Identifiers

Barcodes (UPC/GTIN)
  • 842768014321

Brands