HardyDisk AST Cefiderocol 30¿g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30¿g (FDC30) for in vitro agar diffusion…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2887-2026
- Category
- Medical devices
- Recall date
- 2026-06-08
- Enforcement report date
- 2026-08-12
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Hardy Diagnostics
From the FDA Enforcement Report dated 2026-08-12. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2887-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- HardyDisk AST Cefiderocol 30¿g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30¿g (FDC30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of Enterobacterales, Pseudomonas aeruginosa, Acinetobacter baumannii complex and Stenotrophomonas maltophilia to Cefiderocol,
- Hazard / reason
- Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolonged infection, or worsening clinical outcome.
- Quantity
- 43 units
- Distribution
- U.S. Nationwide distribution in the states of AZ, CA, FL, ID, IN, MA, MS, NJ, NV, NY, OH, PA, and TN.
- Code information
- Catalog Number: Z9435 UD-DI code: Pending Lot Number: 543461
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II