5-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2894-2026
- Category
- Medical devices
- Recall date
- 2026-07-01
- Enforcement report date
- 2026-08-12
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Aniara Diagnostica LLC
From the FDA Enforcement Report dated 2026-08-12. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2894-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.
- Quantity
- 4 units
- Distribution
- US Nationwide distribution in the states of IL, NC, TN.
- Code information
- UDI-DI: 05070002420378, Lot Number: KE132603
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 5070002420378
- Model numbers
- 5D-45905