GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2900-2026
- Category
- Medical devices
- Recall date
- 2026-05-21
- Enforcement report date
- 2026-08-12
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Instrumentation Laboratory
From the FDA Enforcement Report dated 2026-08-12. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2900-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010 00055445011 00055360010.A
- Hazard / reason
- The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
- Quantity
- 109 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.
- Code information
- Part No. 00055360010 UDI-DI 08426950807599 Lots 260505K, 260504AG, 260505F; Part No. 00055360011 UDI-DI 08426950807605 Lots 260505F and 260505J; Part No. 00055407504 UDI-DI 08426950807537 Lot 260505R; Part No. 00055415010 UDI-DI 08426950807698 Lots 260505H and 260505K; Part No. 00055430010 UDI-DI 08426950807759 Lots 260505H, 260505K and 260505F; Part No. 00055430011 UDI-DI 08426950807766 Lot 260505K; Product No. 00055445010 UDI-DI 08426950807810 Lots 260504AA and 260505K; Product No. 00055445011 UDI-DI 08426950807827 Lots 260505F and 260505J; Product No. 00055360010.A UDI-DI 08426950807599 Lot 260504AF;
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 8426950807599
- 8426950807605
- 8426950807537
- 8426950807698
- 8426950807759
- 8426950807766
- 8426950807810
- 8426950807827