Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2903-2026
Category
Medical devices
Recall date
2026-06-15
Enforcement report date
2026-08-12
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
TISSIUM SA

From the FDA Enforcement Report dated 2026-08-12. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Hazard / reason
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Distribution
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Code information
Lot Code: Model No NCOCO2010; UDI-DI 03760279380094; Lot Number 250700042, 251000033, and 251000034
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 3760279380094
Model numbers
  • NCOCO2010

Brands