Medline Micro-Kill Bleach Germicidal Bleach Wipes, 6" x 5", 150 count, Product Number MSC351410AN
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2924-2026
- Category
- Medical devices
- Recall date
- 2026-06-30
- Enforcement report date
- 2026-08-19
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-08-19. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2924-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Medline Micro-Kill Bleach Germicidal Bleach Wipes may exhibit a reduction in bleach concentration to levels below specification prior to the end of the labeled 12-month shelf life. Reduced bleach concentration may compromise the effectiveness of disinfection, potentially allowing for the survival and persistence of pathogenic microorganisms.
- Quantity
- 447,618 units
- Distribution
- US Nationwide distribution in the states and territories of Guam, Puerto Rico.
- Code information
- UDI-DI (each) 10888277652521, UDI-DI (case) 20888277652528, Lot Numbers: All lots produced between 05/13/2025 and 04/21/2026
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II