ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.

Agency
U.S. Food and Drug Administration
Recall number
Z-2937-2026
Category
Medical devices
Recall date
2026-05-18
Enforcement report date
2026-08-19
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Bioporto Diagnostics A/S

From the FDA Enforcement Report dated 2026-08-19. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users.
Quantity
400 units
Distribution
Worldwide - The affected product distribution included the United States and the countries of European Union, Chile, and Hong Kong. No affected product was distributed to Canada; however, Health Canada was notified because the product is registered in Canada.
Code information
Model number: ST101UB / 08787174 190; GTIN code: 5711435000156; Lot-number: 04909125 and 04912125.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 5711435000156
Model numbers
  • ST101UB
Possible model numbers (as listed; not verified as model numbers)
  • 08787174