TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799

Agency
U.S. Food and Drug Administration
Recall number
Z-2960-2026
Category
Medical devices
Recall date
2026-07-08
Enforcement report date
2026-08-26
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Tobii Dynavox Llc

From the FDA Enforcement Report dated 2026-08-26. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
Hazard / reason
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Distribution
US Nationwide distribution.
Code information
UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands