TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2960-2026
- Category
- Medical devices
- Recall date
- 2026-07-08
- Enforcement report date
- 2026-08-26
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Tobii Dynavox Llc
From the FDA Enforcement Report dated 2026-08-26. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2960-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
- Hazard / reason
- Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
- Distribution
- US Nationwide distribution.
- Code information
- UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II