Custom Made Device Implant 6123 Navicular Cage Nominal Right Rev A, Model/Catalog Number 6123N R; foot implant
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2972-2026
- Category
- Medical devices
- Recall date
- 2026-07-10
- Enforcement report date
- 2026-08-26
- Classification
- Class III
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Paragon 28, Inc.
From the FDA Enforcement Report dated 2026-08-26. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2972-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Device was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
- Quantity
- 1 unit
- Distribution
- US Nationwide distribution i the state of MT.
- Code information
- Lot Number AOCIA6123
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class III