Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2983-2026
- Category
- Medical devices
- Recall date
- 2026-07-08
- Enforcement report date
- 2026-08-26
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Fort Defiance Industries, LLC
From the FDA Enforcement Report dated 2026-08-26. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2983-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
- Hazard / reason
- The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
- Distribution
- Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
- Code information
- Lot Code: All serial numbers/ GTIN: (01) 00862992000327
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II