Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2992-2026
Category
Medical devices
Recall date
2026-07-13
Enforcement report date
2026-09-02
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-09-02. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD78018) (Teleflex Product Number/SKU 780-18).
Hazard / reason
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Quantity
145525 units
Distribution
Worldwide distribution - US Nationwide and the country of Panama.
Code information
1) Medline Product Number/SKU HUD78007, UDI/DI 10197344042463 (each) 20197344042460 (case), All lot numbers; 2) Medline Product Number/SKU HUD78009, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 3) Medline Product Number/SKU 780-13, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 4) Medline Product Number/SKU 780-14, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 5) Medline Product Number/SKU 780-15, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 6) Medline Product Number/SKU HUD78018, UDI/DI 10197344010660 (each) 20197344010667 (case), All lot numbers.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 10197344042463
  • 10197344010660

Brands