ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600

Agency
U.S. Food and Drug Administration
Recall number
Z-3034-2026
Category
Medical devices
Recall date
2026-08-06
Enforcement report date
2026-09-02
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
IMRIS Imaging Inc

From the FDA Enforcement Report dated 2026-09-02. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Quantity
13 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.
Code information
114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119; 114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II