STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitative factor…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-3047-2026
- Category
- Medical devices
- Recall date
- 2026-07-29
- Enforcement report date
- 2026-09-09
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Diagnostica Stago, Inc.
From the FDA Enforcement Report dated 2026-09-09. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-3047-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitative factor deficiency test
- Hazard / reason
- A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified
- Quantity
- 5220 units
- Distribution
- US Nationwide distribution in the states and territory of NJ, NY, NJ, VA, PA, FL, AL, GA, TN, OH, IL, MO, NE, TX, CO, UT, CA, PR.
- Code information
- UDI : (01)3607450007466(11)250331(17)270331(10)273074(241)00746, Exp 3/31/2027
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II