STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitative factor…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-3047-2026
Category
Medical devices
Recall date
2026-07-29
Enforcement report date
2026-09-09
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Diagnostica Stago, Inc.

From the FDA Enforcement Report dated 2026-09-09. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitative factor deficiency test
Hazard / reason
A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified
Quantity
5220 units
Distribution
US Nationwide distribution in the states and territory of NJ, NY, NJ, VA, PA, FL, AL, GA, TN, OH, IL, MO, NE, TX, CO, UT, CA, PR.
Code information
UDI : (01)3607450007466(11)250331(17)270331(10)273074(241)00746, Exp 3/31/2027
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands