HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-3064-2026
Category
Medical devices
Recall date
2026-07-31
Enforcement report date
2026-09-09
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Johnson & Johnson Surgical Vision, Inc.

From the FDA Enforcement Report dated 2026-09-09. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU)
Hazard / reason
Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding because forcing the cannula may cause corneal abrasion, iris trauma or descemet's membrane detachment.
Quantity
111,478
Distribution
Worldwide - US Nationwide distribution including in the states of PA, AL, AZ, AR, CA, FL, IL, IN, IA, KY, LA, ME, MI, MN, MS, NJ, NY, NC, OH, OR, TN, TX, WA, WV, HI, CO, MD, VA, WI, ID, NH, SD, OK, MA, NM, DE, GA, SC, CT, NV, MO, RI, KS, ND and the countries of CA, BE, CH, DE, DK, ES, FR GB, HU, IS, IT, NL, NO, PT, SE.
Code information
REF/UDI-DI/Lot: 10300012/5050474755093/UA32232, UA32250, UA32268, UA32290, UA32292, UA32303, UA32371, UP31203, UP31205; 10300016/5050474755116/UA32447, UA32448, UA32449, UB33008, UB33026, UB33082, UB33083, UB33175; 10301012/5050474755109/UB33014, UB33041, UB33057, UB33060, UB33100, UB33190; 10301016/5050474755123/UB33084, UB33194
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands