5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-3084-2026
- Category
- Medical devices
- Recall date
- 2026-08-10
- Enforcement report date
- 2026-09-09
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Fresenius Medical Care Renal Therapies Group, LLC
From the FDA Enforcement Report dated 2026-09-09. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-3084-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- 5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6
- Hazard / reason
- Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
- Quantity
- 120 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
- Code information
- Lots 25KR01227 25NR01257 UDI-DI 00840861102556
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II