Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-3137-2026
- Category
- Medical devices
- Recall date
- 2026-07-23
- Enforcement report date
- 2026-09-23
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-09-23. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-3137-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N
- Hazard / reason
- The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be assured. If a compromised pouch seal is not identified prior to use and the cannula is used, there is a potential risk of infection or other complications associated with the use of a non-sterile device, including hemolysis, foreign body reaction, thromboembolism, and/or organ dysfunction.
- Quantity
- 135 units
- Distribution
- US Nationwide distribution in the state of CA.
- Code information
- UDI-DI: 10198459445941 (each), 40198459445942 (case); Lot Numbers: 25HLA462, 25HLA916, 25JLA847, 25LLA298, 26CLA372
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10198459445941