Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Agency
U.S. Food and Drug Administration
Recall number
Z-3155-2026
Category
Medical devices
Recall date
2026-08-18
Enforcement report date
2026-10-07
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Philips North America Llc

From the FDA Enforcement Report dated 2026-10-07. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
Hazard / reason
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
Quantity
62
Distribution
U.S. and U.S. Territories.
Code information
UDI: (01)00884838101111
Recall type
FDA Mandated
Product type
Devices
Classification (FDA wording)
Class II

Brands