Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;

Agency
U.S. Food and Drug Administration
Recall number
Z-3174-2026
Category
Medical devices
Recall date
2026-08-04
Enforcement report date
2026-09-23
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
PTW FREIBURG

From the FDA Enforcement Report dated 2026-09-23. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;
Hazard / reason
When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.
Quantity
2
Distribution
Worldwide - US Nationwide distribution in the states of GA, NY and the countries of Austria, Belgium, Switzerland, Czechia, Germany, Spain, France, Netherlands, Norway, Poland, Portugal, Romania, T¿rkiye, Chile, United Arab Emirates, Saudi Arabia, Mexico, Hong Kong, Brazil, Tajikistan, Ecuador, Iran, Thailand, Serbia, South Africa, Kazakhstan, Algeria.
Code information
Model/Catalog Number: S070031.520; UDI-DI: +EPTWS070031Q5200; Software Version 3.1.18;