TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-3204-2026
- Category
- Medical devices
- Recall date
- 2026-07-30
- Enforcement report date
- 2026-09-23
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Howmedica Osteonics Corp.
From the FDA Enforcement Report dated 2026-09-23. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-3204-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
- Hazard / reason
- Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
- Quantity
- 12,156 US
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
- Code information
- 620-00-28D UDI-DI 07613327313581 Lots MHANP2 MHD1N0
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 7613327313581