SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE BIPOLAR 40MM 6074-4022 SYSTEM ONE BIPOLAR 43MM 6074-4322 SYSTEM ONE…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-3205-2026
Category
Medical devices
Recall date
2026-07-30
Enforcement report date
2026-09-23
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Howmedica Osteonics Corp.

From the FDA Enforcement Report dated 2026-09-23. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE BIPOLAR 40MM 6074-4022 SYSTEM ONE BIPOLAR 43MM 6074-4322 SYSTEM ONE BIPOLAR 44MM 6074-4422 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Hazard / reason
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Quantity
12,156 US
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Code information
78-5-0910 UDI-DI 07613327366006 Lot 30582401 6074-4022 UDI-DI 07613327031775 Lots 2M5NTY LH25R5 M36N83 6074-4322 UDI-DI 07613327031805 Lot W40E3T 6074-4422 UDI-DI 07613327031812 Lot R84K20
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 7613327366006
  • 7613327031775
  • 7613327031805
  • 7613327031812

Brands