CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-3212-2026
Category
Medical devices
Recall date
2026-08-20
Enforcement report date
2026-09-23
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
ICU Medical, Inc.

From the FDA Enforcement Report dated 2026-09-23. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product Codes: 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24JP 21-7600-24JP 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24
Hazard / reason
Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.
Quantity
185,688 units
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, LA, MI, MO, MS, TN, TX, VA, WI and the countries of Austria, Australia, Belgium, Costa Rica, Germany, Estonia, Spain, Finland, France, Great Britain, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, and Sweden.
Code information
21-7600-24 UDI-DI 30610586044001 Lots 6026878 6037784 6054015 6062724 6070616 6082332 6092886 6101906 6108556 6126029 6140893 6143454 6147696 6147697 6162769 6155095 6166032 6171587 6171586 21-7609-24 UDI-DI 30610586044018 Lot 6022079 6037697 6037698 6053978 6061525 6072970 6072964 6070575 6077773 6092846 6101798 6108514 6108515 6108516 6140862 6140861 6140865 6155132
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 30610586044001
  • 30610586044018

Brands