VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

Agency
U.S. Food and Drug Administration
Recall number
Z-3231-2026
Category
Medical devices
Recall date
2026-07-24
Enforcement report date
2026-09-30
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Biomerieux Inc

From the FDA Enforcement Report dated 2026-09-30. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.
Quantity
97680 units
Distribution
International distribution to the countries of Austria, Germany, Luxembourg.
Code information
UDI-DI 03573026637712; Lot Number 1333393504
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 3573026637712

Brands