Ingenia Elition S and Ingenia Elition X; Model Numbers (REF): 1. 781357(S), 2. 782106(S), 3. 782137(S), 4. 781358(X)…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-3239-2026
Category
Medical devices
Recall date
2026-08-21
Enforcement report date
2026-09-30
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Philips North America Llc

From the FDA Enforcement Report dated 2026-09-30. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Ingenia Elition S and Ingenia Elition X; Model Numbers (REF): 1. 781357(S), 2. 782106(S), 3. 782137(S), 4. 781358(X), 5. 782107(X), 6. 782119(X), 7. 782136(X), 8. 782158(X);
Hazard / reason
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Quantity
596
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia;
Code information
1. Model Numbers (REF): 781357(S); UDI-DI: 884838088108; Serial Numbers: 46083, 46045, 46095, 46119, 46121, 46024, 46044, 46055, 46061, 46052, 46001, 46005, 46011, 46048, 46057, 46060, 46066, 46078, 46081, 46016, 46026, 46032, 46033, 46037, 46046, 46054, 46058, 46059, 46068, 46069, 46075, 46082, 46003, 46019, 46111, 46007, 46107, 46067, 46093, 46030, 46031, 46049, 46050, 46089, 46014, 46065, 46071, 46077, 46088, 46109; 2. Model Numbers (REF): 782106(S); UDI-DI: 884838098329; Serial Numbers: 46247, 46250, 46316, 46376, 46200, 46368, 46324, 46331, 46338, 46363, 46367, 46383, 46394, 46398, 46399, 46400, 46201, 46202, 46220, 46225, 46352, 46310, 46234, 46205, 46207, 46209, 46211, 46214, 46215, 46226, 46228, 46235, 46237, 46239, 46240, 46241, 46282, 46281, 46297, 46304, 46317, 46354, 46212, 61010, 46206, 46362, 46393, 46348, 46217, 46289, 46233, 46353, 46213, 46218, 46222, 46384, 46224, 46265, 46286, 46390, 46263, 46307, 46387, 46403; 3. Model Numbers (REF): 782137(S); UDI-DI: 884838108615; Serial Numbers: 64004, 64008, 64024, 64028, 64031, 64046, 64047, 64002, 64003, 64009, 64013, 64034, 64045, 64050, 64030, 64029, 64049, 64051; 4. Model Numbers (REF): 781358(X); UDI-DI: 884838088115; Serial Numbers: 45107, 45142, 45145, 45216, 45241, 45284, 45121, 45257, 45154, 45193, 45253, 45106, 45131, 45259, 45298, 45300, 45302, 45313, 45330, 45344, 45345, 45349, 45352, 45359, 45360, 45363, 45367, 45368, 45370, 45371, 45372, 45377, 45378, 45381, 45390, 45391, 45394, 45396, 45397, 45409,…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 884838088108
  • 884838098329
  • 884838108615
  • 884838088115
  • 884838098336
  • 884838104129
  • 884838108608
  • 884838115699

Brands