Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-3257-2026
Category
Medical devices
Recall date
2026-08-03
Enforcement report date
2026-09-30
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Baxter Healthcare Corporation

From the FDA Enforcement Report dated 2026-09-30. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No
Hazard / reason
Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.
Quantity
2
Distribution
US Nationwide distribution in the states of Colorado and Florida.
Code information
Lot Code: UDI: 00887761981690 Serial Numbers: W078SY3553 and Z309SY7853
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 887761981690

Brands