Fisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + Urine Cassette Test…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-3274-2026
- Category
- Medical devices
- Recall date
- 2026-08-20
- Enforcement report date
- 2026-10-07
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Alere San Diego, Inc.
From the FDA Enforcement Report dated 2026-10-07. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-3274-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Fisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + Urine Cassette Test, Model/ Part: FHC-102/9004074, FHC-102/9004073; Alere hCG Cassette (25mIU/mL) in Urine, Model/ Part: FHC-102/92217; Cardinal Health hCG Cassette Rapid test CLIA Waived, Model/ Part: FHC-102/B1077-22; Medline HCG Pregnancy Test Cassette, Model/ Part: FHC-102/MPH12025HCG; Pro Advantage By NDC Urine Pregnancy Cassette Device, Model/ Part: FHC-102/P080063;
- Hazard / reason
- hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
- Quantity
- 18,139,960
- Distribution
- Worldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Ireland, Spain, Portugal, New Zealand, Hong Kong, Australia, Papua New Guinea, Fiji, Guam, Sweden.
- Code information
- All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 23900526/00613647000526/1155811105, 2018-01-15; 9004074/10304040024116/1155875704, 2014-06-09; 9004073/10304040024109/1155875704, 2014-06-09; 92217/10815845020109/1156015703, 07/16; B1077-22/20885380020251/1155904704, 9/2014; MPH12025HCG/10888277191877/1156093302, 2015-03-20; P080063/M220P0800633/40001853, 2023-03-30
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II