Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 W Model/Catalog…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-3327-2026
Category
Medical devices
Recall date
2026-07-17
Enforcement report date
2026-10-07
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
SPECTRUM MEDICAL SRL

From the FDA Enforcement Report dated 2026-10-07. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 W Model/Catalog Number: HX55V-S0W Software Version: N/A Product Description: The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 8 l/min. Quantum PureFlow Standard Heat Exchanger is a single use device which allows heating/cooling the blood during extracorporeal circulation. The device is sold sterile in a blister pack in a multiple box of 4 units. Device surfaces in contact with blood are treated with a phosphorylcholine-based biocompatible coating. The device has been designed to be powered by heaters coolers that use water or a glycol-based heat transfer fluid (HTF). Component: N/A
Hazard / reason
Due to blood-to-glycol leak in the heat exchanger.
Quantity
9448
Distribution
US Nationwide distribution in the states of AL, CA, FL, KS MA, MD, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WI.
Code information
Lot Code: 18051160302205 (HX55V-S0W)
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands