- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-3330-2026
- Category
- Medical devices
- Recall date
- 2026-07-16
- Enforcement report date
- 2026-10-07
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Tandem Diabetes Care, Inc.
From the FDA Enforcement Report dated 2026-10-07. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-3330-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No
- Hazard / reason
- A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
- Quantity
- 177
- Distribution
- US Nationwide distribution.
- Code information
- Lot Code: UDI-DI: 00389152470641 Software Version: 2.13 (iOS only)
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 389152470740
- 389152470641