Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 /…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-3332-2026
Category
Medical devices
Recall date
2026-08-03
Enforcement report date
2026-10-07
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Beckman Coulter Mishima K.K.

From the FDA Enforcement Report dated 2026-10-07. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only.
Hazard / reason
Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.
Quantity
492 units
Distribution
Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.
Code information
Products containing affected software: Item Numbers / UDI-DI codes: Item Number: B96692 UDI-DI code: 14987666542965 Item Number: B96693 UDI-DI code: 14987666542972 Item Number: B12183 UDI-DI code: 14987666540671 Item Number: N3151400 UDI-DI code: Due to the age of the device and the fact that it has exceeded its storage period, the UDI number is not available.)
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 14987666542965
  • 14987666542972
  • 14987666540671

Brands